20-L-Apparatus

for the determination of explosion indices of dust, gas and hybrid mixtures
KSEP For the efficient prevention of dust explosions by preventive explosion protection (e.g. explosion venting, explosion suppression) it is necessary to know the optimum explosion indices: 

- the dust explosibility
- the lower explosibility limit (LEL)
- the maximum explosion pressure (Pmax)
- the dust-specific characteristic (Kmax)
- the limiting oxygen concentration (LOC)

These characteristics, especially those that describe the explosion behavior of combustible dusts, can be determined according to a standard procedure in the 20-l laboratory apparatus.

 

hw_ksep.jpg

The test chamber of the laboratory apparatus is a hollow sphere made of stainless steel. A water jacket serves to dissipate the heat of the explosions and to maintain a thermostatically controlled test temperature.
 

Free Download:

KSEP_ICO.gif

Instruction: download and start "Ksep_new.exe" (4.5 MB). 


K_PDF.gif


KSEP-Manual

german version

english version

     
KSEP-Software, Version 6.0 f

- 32-bit software
- support of larger equipments e.g. 1m3-vessel
- automatic check of the explosion indices Pmax and Kmax
- this software is running under Windows 95 / 98 / NT / 2000 / ME / XP / Vista
- Help-files in English and German language are included.